Manufacturing Companies for Solid Waste Treatment Oxygen Analyzer - Nernst N2001 oxygen analyzer – Litong

Short Description:


Product Detail

Product Tags

Related Video

Feedback (2)

To create much more price for clients is our company philosophy; purchaser growing is our working chase for Water Vapor Analyzer Price, Zirconia Oxygen Analyser, Handheld Gas Analyzer, On account of superior quality and competitive price , we will be the market leader, please don’t hesitate to contact us by phone or email, if you are interested in any of our products.
Manufacturing Companies for Solid Waste Treatment Oxygen Analyzer - Nernst N2001 oxygen analyzer – Litong Detail:

Application range

The Nernst N2001 oxygen analyzer is widely used in the detection of oxygen content in the combustion process of electric power, metallurgy, steel, chemical industry, ceramics, incineration, etc.It can also directly monitor the oxygen content in the flue gas of boilers, kilns, sintering furnaces, heating furnaces, pit annealing furnaces, etc. during or after combustion.During the measurement, the temperature in the furnace and the flue can be from normal temperature to 1400°C high temperature.The oxygen analyzer provides real-time oxygen content O2% (percentage) parameters and oxygen potential millivolt values in the furnace or flue.

Technical characteristics

 Input function: One oxygen analyzer can be connected to a oxygen probe to display the measured oxygen content in real time.

Multi-channel output control: the analyzer has one 4-20mA current output and network communication interface RS485.

 Measurement range: The oxygen measurement range is 0 to 100% oxygen.

Alarm setting: The analyzer has 1 general alarm output and 3 programmable alarm outputs.

 Automatic calibration: The analyzer has an automatic calibration function, and the user can customize the calibration time and number of calibrations according to their needs.

• Automatic dust cleaning: The analyzer has the function of automatically cleaning the probe. The user can clean the probe regularly according to the needs, eliminating the need for manual on-site dust cleaning.

Intelligent system: The analyzer can complete the functions of various settings according to the predetermined settings.

• Display output function: The analyzer can display the date, current oxygen content, probe temperature, current oxygen millivolt value and 14 first-level status displays and 11 second-level status displays.

 Safety function: When the furnace is out of use, the user can control to turn off the heater of the probe to ensure safety during use.

 Installation is simple and easy: the installation of the analyzer is very simple and there is a special cable to connect with the zirconia probe.

Specifications

Inputs

One zirconia oxygen probes or sensors

Outputs

Linear 4~20mA DC

Display mode

128×64 dot liquid crystal display

Probe heating method

PID control

Dust cleaning and standard gas calibration

The analyzer has an automatic calibration function and an automatic dust cleaning function.It can also automatically control the solenoid valve switch.

ary Parameter Display

AlarmsParameter Display

High and low oxygen alarms can be set arbitrarily.

AccuracyP

± 1% of the actual oxygen reading with a repeatability of 0.5%.

Reaction rateP

Indirect heating measurement is about 3 seconds

Direct heating in 30 seconds

Core detection reaction speed: 0.0001s

Range of Local Indication

0 to 100% Oxygen

Serial/Network Interface

RS232

RS485 MODBUSTM

Reference gas

There is a miniature brushless motor vibration pump in the analyzer for reference gas.

Power Ruireqements

85VAC to 264VAC 3A

Operating Temperature

Operating Temperature -25°C to 55°C

Relative Humidity 5% to 95% (non-condensing)

Degree of Protection

IP65

IP54 with internal reference air pump

Dimensions and Weight

300mm W x 180mm H x 100mm D 2.5kg


Product detail pictures:

Manufacturing Companies for Solid Waste Treatment Oxygen Analyzer - Nernst N2001 oxygen analyzer – Litong detail pictures


Related Product Guide:

We insist over the principle of enhancement of 'High high quality, Efficiency, Sincerity and Down-to-earth working approach' to offer you with superb assistance of processing for Manufacturing Companies for Solid Waste Treatment Oxygen Analyzer - Nernst N2001 oxygen analyzer – Litong, The product will supply to all over the world, such as: Islamabad, Netherlands, Hongkong, With the goal of "zero defect". To care for the environment, and social returns, care employee social responsibility as own duty. We welcome friends from all over the world to visit and guide us so that we can achieve the win-win goal together.
  • Customer service staff and sales man are very patience and they all good at English, product's arrival is also very timely, a good supplier.
    5 Stars By Amber from Bulgaria - 2017.09.29 11:19
    This company has the idea of "better quality, lower processing costs, prices are more reasonable", so they have competitive product quality and price, that's the main reason we chose to cooperate.
    5 Stars By Alice from Cambodia - 2017.11.11 11:41

    Related Products